3. Press the MENU key once to activate the menu items tab.
Press the MENU key multiple times to tab right and the
ESC key multiple times to tab left until the PRB menu is
highlighted and displaying the submenu items.
4. Press the UP and DOWN arrow keys to scroll through the
sub menu items until ZERO PROBE is highlighted.
5. Press the ENTER key to display the confirmation screen.
6. If a coating transducer was identified use the UP and
DOWN arrow keys to toggle coating on/off.
7. Press the OK key to complete the probe zero function, or
ESC key to cancel the probe zero function.
8. Remove the transducer from the probe zero disk, and
proceed to the calibration section.
NOTE: The value that is displayed will change depending on
the current velocity setting in the TI-CMX. Disregard the number
that is displayed. It is not important. What is important is
accurately performing the steps outlined above to insure
reliability of the probe zero calculation.
One Point Material Calibration
For the purposes of this quick start section, we’ll only be covering the most common
one point calibration option to determine the sound velocity of the test material. It would
be very handy to carry a set of mechanical calipers to use in conjunction with the
TI-CMX for calibration in the field:
NOTE: Be sure that the probe zero procedure has been
performed prior to performing this calibration procedure.
1. Physically measure an exact sample of the material or a
location directly on the material to be measured using a
set of calipers or a digital micrometer.
2. Apply a drop of couplant on the transducer and place the
transducer in steady contact with the sample or actual test
material. Be sure that the reading is stable and the repeat-
ability indicator, in the top left corner of the display, is fully
lit and stable. Press the MENU key once to activate the menu
items tab. Press the MENU key multiple times to tab right
and the ESC key multiple times to tab left until the CAL
menu is highlighted and displaying the submenu items.
3. Use the UP and DOWN arrow keys to scroll through the
sub menu items until MATL 1PT is highlighted.
4. Press the ENTER key to display the Digits Edit Box.
5. Press the UP and DOWN arrow keys to scroll the highlighted
value.
6. Press the LEFT and RIGHT arrow keys to scroll the digit
locations.
-7-
Coating Probe
Identified
Coating Probe
Not Identified
12.0 USING THE UTILITY SOFTWARE
12.1 Computer System Requirements
Datacomm will run on many different operating systems: Windows 98 (1st or 2nd
edition), Windows NT 4.0 with Service Pack 5, Windows ME, Windows XP, Windows
2000 Professional, Windows 2000 Server, or Windows 2000 Advanced Server operating
systems running on Intel or AMD hardware.
A Pentium 166MHz or faster processor with at least 32 megabytes of physical RAM is
required. You should have 40 megabytes of free disk space before attempting to install
Datacomm.
Datacomm requires an available communications port in order to transfer data to and
from the TI-CMX. Datacomm supports COM1, COM2, COM3, and COM4.
USB to Serial Converter
Some newer laptop computers do not have standard serial ports. In this case it is possi-
ble to use a USB to Serial converter. If a serial to USB cable is needed, contact
Electromatic
12.2 Installing Datacomm
Datacomm comes on a CD-ROM with an automatic installer program. Place the CD in
your computer’s CD tray and close the door. Open the CD-ROM by double clicking on
the My Computer ICON, then double click on the CD. Finally, double click on the
SETUP icon to begin the installation.
Refer to the help section in Datacomm software for the complete operating manual,
setup, and operation.
12.3 Using the XFER menu (TI-CMX)
The XFER menu of the TI-CMX will be used in conjunction with
the Datacomm PC software. The steps below outline the procedure
for accessing the XFER menu and basic operation as follows:
1. Press the MENU key once to activate the menu items tab. Press
the MENU key multiple times to tab right, and the ESC key
multiple times to tab left, until the XFER menu is highlighted
and displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll through the
sub menu items until the desire option is highlighted.
3. Press the ENTER key to activate the option selected.
4. Once the Backup or Restore function has been completed, press the MEAS key once
to return to the menu items, or twice to return to the measurement mode.