– 8 –
5.0 Damping Mode
The KXE is equipped with an electronic damping which ensures steady readings when
tension fluctuates. This is achieved by averaging the measured values at the set update
rate.
NOTE: Before switching on the damping mode, it is recommended that you measure
the first values without damping enabled.
5.1 To switch damping on and off:
1. Insert process material and take a mesurement as directed in Section 4.0
2. Press the DAMP key. The display shows the set damping factor.
3. Release the DAMP key.
The display shows DAMP below the currently measured value.
0000฀
3. To switch off damping, press the DAMP key again.
The display shows only the currently measured value.
0000฀
5.2 Modifying the Dampling Factor
The tension meter is factory preset to a damping factor of 12. The tension values are
averaged for the display in the following way:
12 old values + 4 new values
16
The damping factor can be modified in 15 steps from 01 = low damping:
1 old value + 15 new values
16
to 15 = high damping:
15 old values + 1 new value
16
Damp
– 21 –
3. The DISPLAY blinks, showing
22
the average value (AVG) of the first
measuring period (POS: 1) and the
symbol.
4. Press the RECALL key. The DISPLAY blinks, showing
23
the
last value (LAST) of the measuring period.
5. Press the RECALL key. The DISPLAY blinks, showing
24
the
maximum value (MAX) of the measuring period and the
symbol.
6. Press the RECALL key. The DISPLAY blinks, showing
21
the
minimum value (MIN) of the measuring period and the
symbol.
7. Press the RECALL key. The DISPLAY blinks, showing
26
Peak
the
maximum peak value (PEAK MAX) of the measuring period, the PEAK indicator
and the
symbol.
8. Press the RECALL key. The DISPLAY blinks,
19
Peak
showing the
minimum peak value (PEAK MIN) of the measuring period, the PEAK indicator
and the
symbol.
9. Press the RECALL key. The measured values no. 1 — 10 of the first measuring
period can be recalled.
10. Press the RECALL key. The next measuring period (POS: 2) is shown on the
DISPLAY and the sequence is repeated, starting with the average value (AVG).